Recent Announcements
Instrument Status Post SSD2023
Posted Sep 09, 2023
LBC
During SSD, the filters and accessible lenses were cleaned, and the TMS collimators were reinstalled and realigned. A software service was put in place to simplify operation of the LBCs with TMS.
During restart, the LBCs were …
>> Read More
PEPSI upgrades and work status SSD23
Posted Sep 01, 2023
All is back to normal and PEPSI is ready for observing in 23B after extensive work during the summer of 2023.
The two new STA1600 CCDs were integrated into the existing dewars at STA premises in California over the August …
>> Read More
Resolved: LUCI2 Detector Stage Failure – LUCI2 Unavailable
Posted Jun 16, 2023
Resolved during SSD23: LUCI2 had a failure of its detector focus stage on June 14, 2023. Repeated attempts at restoring function have been unsuccessful. The instrument is currently offline for N1.8 & N3.75 (seeing limited and ESM) use as they …
>> Read More
SX M2 swap
Posted Jun 03, 2023
During the afternoon of May 25th, the SX M2 adaptive secondary (ASM) was removed and the rigid secondary installed in its place. The coating on the SX ASM had been degrading, and it is being sent to Italy for recoating. …
>> Read More
Restricted LUCI1-AGw1 Guider Patrol Field (resolved)
Posted Mar 04, 2023
Update Jan 10 2024: The issue was resolved during SSD23 and the AGw patrol field is back to the nominal area described in the LUCI pages (here). The problem was traced to the camera cable, where a couple of wires …
>> Read More
Resolved: PEPSI Blue CD Position Issue
Posted Feb 10, 2023
Resolved with PEPSI SSD2023 upgrade and work.
Update 2023-02-10
The blue CD is degrading its operational performance: now it cannot be re-positioned from UMAC at all. It can no longer move down to CD1. It only can traverse between CD2 …
>> Read More